06 | Keysight | B1500A Semiconductor Device Analyzer - Data Sheet
Maximum module configuration
The total power consumption of all SMU modules cannot exceed 84 W. Under
this rule, the B1500A can contain any combination of the following SMUs:
–– Up to 10 MPSMUs
–– Up to 10 HRSMUs
–– Up to 4 HPSMUs
–– Up to 4 MCSMUs
Only one single-slot MFCMU can be installed per B1500A mainframe. Up to five
single-slot HV-SPGUs can be installed per mainframe. Up to five single-slot
WGFMUs can be installed per mainframe.
When one or more WGFMU modules are installed in the B1500A mainframe,
the following table applies. Multiply the values given below by the number of
installed modules of that type and add the products together. The sum of the
products must be less than or equal to 59 for the configuration to be permissible.
MPSMU
2
HRSMU
2
HPSMU
14
MCSMU
5
MFCMU
7
HV-SPGU
12
WGFMU
10
Maximum voltage between common and ground
≤ ±42 V
Ground unit (GNDU) specification
The GNDU is furnished standard with the B1500A mainframe.
Output voltage: 0 V ±100 µV
Maximum sink current: ±4.2 A
Output terminal/connection: Triaxial connector, Kelvin (remote sensing)
GNDU supplemental characteristics
Load capacitance: 1 µF
Cable resistance:
–– For IS ≤ 1.6 A: force line R < 1 Ω
–– For 1.6 A < IS ≤ 2.0 A: force line R < 0.7 Ω
–– For 2.0 A < IS ≤ 4.2 A: force line R < 0.35 Ω
–– For all cases: sense line R ≤ 10 Ω
–– Where IS is the current being sunk by the GNDU.
Peripherals and interface
Data storage
Hard disk drive, DVD-R drive
Interfaces
GPIB, interlock, USB (USB 2.0, front 2, rear 2),
LAN (1000BASE-T/100BASE-TX/10BASE-T),
trigger in/out, digital I/O, VGA video output
Remote control capabilities
–– FLEX commands (GPIB)
–– EasyEXPERT remote control function
(LAN)
Trigger I/O
–– Only available using GPIB FLEX
commands.
–– Trigger in/out synchronization pulses
before and after setting and measuring
DC voltage and current. Arbitrary trigger
events can be masked or activated
independently.
Furnished accessories
–– Keyboard
–– Mouse
–– Stylus pen
–– Power cable
–– Manual CD-ROM
–– EasyEXPERT CD-ROM
–– Software CD-ROM
(including VXI plug&play
driver and utility tools)
–– Software entitlement document
for EasyEXPERT
Furnished software
–– EasyEXPERT group+
–– VXI plug&play driver for the B1500A
–– MDM file converter
This tool can convert the EasyEXPERT
file (XTR/ZTR) to Keysight IC-CAP
MDM file format.
The EasyEXPERT file of the following
measurements performed in the classic
mode is only supported:
–– IV Sweep
–– Multi channel IV Sweep
–– CV Sweep
–– 4155/56 setup file converter tool
This tool can convert 4155 and 4156
measurement setup files (file extensions
MES or DAT) into equivalent EasyEXPERT
classic test mode setup files.