Specification Conditions
This document contains specifications and supplemental information for the 2606B System SourceMeter® instrument. Specifications are
the standards against which the 2606B instruments are tested. Upon leaving the factory, the 2606B instruments meet these specifications.
Supplemental and typical values are nonwarranted, apply at 23°C, and are provided solely as useful information. Specifications are for
individual modules.
Source and measurement accuracies are specified at the 2606B terminals under these conditions:
1. 23°C ±5°C, <70 percent relative humidity
2. After a two-hour warm-up period
3. Speed normal (1 NPLC)
4. A/D autozero enabled
5. Remote sense operation or properly zeroed local operation
6. Calibration period: One year
DC Power Specifications
Maximum output power
and source limits 1, 2
Voltage
20.2 W maximum
20.2 V at 1.0 A, –20.2 V at –1.0 A
6.06 V at 3.0 A, –6.06 V at –3.0 A
+3 A
Current
20.2 W maximum
1.01 A at 20 V, –1.01 A at –20 V
3.03 A at 6 V, –3.03 A at –6 V
+1 A
0A
DC & Pulse
–1 A
–3 A
–20 V
–6 V
0 V +6 V
+20 V
Voltage Accuracy Specifications 3, 4
Range
100 mV
1V
6V
20 V
Programming
Resolution
5 µV
50 µV
50 µV
500 µV
Source
Accuracy
±(% reading + volts)
0.02% + 250 µV
0.02% + 400 µV
0.02% + 1.8 mV
0.02% + 12 mV
Typical Noise (peak to peak)
0.1 Hz to 10 Hz
20 µV
50 µV
100 µV
500 µV
Display
Resolution
100 nV
1 µV
1 µV
10 µV
Measure
Accuracy
±(% reading + volts)
0.015% + 150 µV
0.015% + 200 µV
0.015% + 1 mV
0.015% + 8 mV
Notes
1. For additional power derating information for various load and temperature conditions, refer to “Operating Boundaries” in the Series 2606B Reference Manual.
2. The System SourceMeter® will allow quadrant II and quadrant IV operation, however, it is intended for low power or short transient behavior. Operation in quadrant II and quadrant IV could
result in an overtemperature error.
3. For temperatures 0°C to 18°C and 28°C to 50°C, accuracy is degraded by ±(0.15 × accuracy specification)/°C.
4. Add 50 μV to source accuracy specifications per volt of HI lead drop.