High-Speed Response for Capturing Transient Events
- Function Details -
Large memory capacity allows long-term
recording of high- speed data
In the standard configuration, the 8841 and 8842 can store a
total of 8 mega-words, and with memory expansion 32 mega-
words, using internal solid-state memory. This provides ample
capacity to store data for all 16 channels. The table at right
shows possible recording times, according to the time axis
setting and the number of channels in use. A reduction in the
number of channels prolongs the recording time.
* The table applies to the standard memory configuration. When the optional
MEMORY BOARD 9608 is installed, recording times are extended by a
factor of 4 (from 16 mega-words/channel, 160,000 divisions for 2 channels to
2 mega-words/channel, 20,000 divisions for 16 channels).
Memory segmentation function
When using the memory recorder function, the data memory can
be divided into a maximum of 255 blocks. Data can be written
sequentially to the memory blocks, and the waveform in a reference
block and any other block can be superimposed and compared.
* In the standard memory configuration (8 mega-words), the maximum
number of blocks using the 16 channel setting is 127. When the memory
expansion is installed, or when the channel setting is 2 to 8, the maximum
number of blocks is 255.
Recording length using 16 channels
Segmentation into 3 blocks
Segmentation
Segmentation into 127 blocks
A/D
conversion
Input voltage
Insulation
Function outline of memory recorder
Write on memory
Thermal printer
A/D
conversion
Insulation
Input voltage
Input waveform
Sampling period
Time
Output waveform
Time axis
100μs/DIV
200μs/DIV
500μs/DIV
1ms/DIV
2ms/DIV
5ms/DIV
10ms/DIV
20ms/DIV
50ms/DIV
100ms/DIV
200ms/DIV
500ms/DIV
1s/DIV
2s/DIV
5s/DIV
10s/DIV
30s/DIV
1min/DIV
2min/DIV
5min/DIV
Sampling
period
1μs
2μs
5μs
10μs
20μs
50μs
100μs
200μs
500μs
1ms
2ms
5ms
10ms
20ms
50ms
100ms
300ms
0.6s
1.2s
3.0s
2-channel setting
4 MW/channel 40,000 DIV
4s
8s
20 s
40 s
1 m 20 s
3 m 20 s
6 m 40 s
13 m 20 s
33 m 20 s
1 h 6 m 40 s
2 h 13 m 20 s
5 h 33 m 20 s
11 h 6 m 40 s
22 h 13 m 20 s
2 days 7 h 33 m 20 s
4 days 15 h 6 m 40 s
13 days 21 h 20 m
27 days 18 h 40 m
55 days 13 h 20 m
138 days 21 h 20 m
16-channel setting
500 kW/channel 5,000 DIV
0.5 s
1s
2.5 s
5s
10 s
25 s
50 s
1 m 40 s
4 m 10 s
8 m 20 s
16 m 40 s
41 m 40 s
1 h 23 m 20 s
2 h 46 m 40 s
6 h 56 m 40 s
13 h 53 m 20 s
1 day 17 h 40 m
3 days 11 h 20 m
6 days 22 h 40 m
17 days 8 h 40 m
Zoom function *In memory recorder function
To make the most of the large-capacity memory, it is possible
to display a compressed waveform simultaneously with a
magnified waveform. Since the 8841 and 8842 are capable
of storing a large amount of data, high-speed sampling is also
possible for waveforms with a long duration. Accordingly,
while observing the compressed image of the entire waveform,
it is also possible to observe the magnified details of desired
parts. Compressed display of a part of the entire waveform is
also possible.
Clock input for external sampling *In memory recorder function
The sampling rate for the memory recorder can be controlled by
the timing of an external clock signal. This is useful for example to
collect data synchronized to the running cycle of an engine.
Sampling controlled by external
clock signal
Input waveform
Maximum frequency 500kHz
Sampling period
Time